Instructions for use Carina SW 3.2n 133
Technical data
Settings for typical ventilation
Factory-set ventilation parameters
Ventilation mode PC-AC
VT 500 mL
Pinsp 15 mbar (or hPa or cm H
2O)
PEEP 3 mbar (or hPa or cm H
2O)
FiO
2 40 %
RR 15/min
Ti 1.3 s
Timax 4 s
I:E 1:2
ΔPsupp 5 mbar (or hPa or cm H
2O)
Ramp Auto
Pmax 15 mbar (or hPa or cm H
2O)
VG Off
Trigger Normal
Apnea ventilation On
RRapn 15/min
VTapn 500 mL
Anti Air Shower On
Night mode Off
Aut.Adapt Off
Ventilation mode PC-BIPAP
Pinsp 15 mbar (or hPa or cm H
2O)
PEEP 5 mbar (or hPa or cm H
2O)
RR 12/min
Ramp Auto
I:E 1 : 1
FiO
2 40 %
O
2 suppl. HPO
Operating temperature 25 °C (77 °F)