Instructions for use Carina SW 3.2n 31
Operating concept
Setting ventilation modes
1 Press (A) for the ventilation mode. The color
changes to yellow.
2 To set the ventilation mode turn the rotary knob
(B).
3 To confirm the ventilation mode push the rotary
knob (B). The color of the key changes to dark
green.
Setting ventilation parameters
1 Set ventilation mode.
2 Press (C) for the applicable ventilation parame-
ter. The color changes to yellow.
3 To set the value turn the rotary knob (B).
4 To confirm the value push the rotary knob (B).
The color of the key changes to dark green.
Exceeding the set limit of a ventilation
parameter
If a setting limit of a ventilation parameter is
reached, the device displays a message in the in-
formation line (A).
1 Set the ventilation parameter named in the
message and confirm with the rotary knob.
2 The setting limit of the ventilation parameter can
now be exceeded. Set ventilation parameter ac-
cordingly and confirm with the rotary knob.
Display of associated ventilation param-
eters
If a ventilation parameter (B) is set, the associated
ventilation values are shown in the display (A). The
display is closed after confirmation of the setting.
080080
108115