Instructions for use Carina SW 3.2n 59
Getting started
Checking readiness for operation
After the device is switched on or after the breath-
ing circuit is changed, the operational readiness of
the device must be verified by means of the device
check.
The checklist on pages 60 and 61 can be copied for
documentation purposes.
Check the test steps in the copied checklist
once they have been successfully completed.
The following functions are tested:
LEDs, displays and alarm tone
Upper alarm limit Paw
Lower alarm limit Paw
Apnea ventilation
Connecting the test lung
Use only the test lungs specified in the list of acces-
sories.
Connect the appropriate test lung to the breath-
ing circuit.
CAUTION
Do not use the device unless the device check
was successful.