Technical data
122 Instructions for use Carina SW 3.2n
Ambient conditions
During operation
Temperature 5 to 35 °C (41 to 95 °F)
Ambient pressure 900 to 1100 hPa (13.1 to 16 psi)
From 700 to 900 hPa (10.2 to 13.1 psi), reduced
functionality in terms of maximum flow and pres-
sure delivery
Relative humidity 5 to 95 %, non-condensing
Altitude Max. 3000 m (9842 ft)
During storage and transport
Temperature (up to 1 day) –20 to 60 °C (–4 to 140 °F)
Temperature (up to 1 month) –20 to 55 °C (–4 to 131 °F)
Temperature (up to 3 months) –20 to 35 °C (–4 to 95 °F)
Ambient pressure 500 to 1100 hPa (7.3 to 16 psi)
Relative humidity 5 to 95 %, non-condensing
The ambient conditions may be further restricted,
depending on the accessories used. Observe cor-
responding instructions for use.
Setting values
Ventilation systems Breathing circuit with leakage valve
Breathing circuit with controlled expiratory valve
Ventilation modes VC-SIMV, VC-AC, PC-BIPAP, PC-AC,
SPN-CPAP/PS (VG)
Tidal volume VT
Range 100 to 2000 mL, BTPS
Resolution 10 mL
Positive end-expiratory pressure PEEP
Range 3 to 20 mbar (or hPa or cm H
2O)
Resolution 1 mbar (or hPa or cm H
2O)
Pressure limitation Pmax
Range (leakage valve) 5 to 40 mbar (or hPa or cm H
2O)
Range (expiratory valve) 5 to 50 mbar (or hPa or cm H
2O)
Resolution 1 mbar (or hPa or cm H
2O)