Technical data
130 Instructions for use Carina SW 3.2n
Apnea alarm time Tapn
Alarm If no breathing activity has been detected
Range 5 to 60 s, Off
Resolution 1 s
Operating data characteristics
Power supply
Mains power 100 V to 240 V AC, 50/60 Hz
Power consumption At 100 V AC max. 1.1 A
At 240 V AC max. 0.5 A
Maximum power consumption (operation,
charging internal battery)
90 W
Typical power consumption (operation, not
charging internal battery)
35 W
Internal battery
Type NiMH enclosed, maintenance-free
Nominal capacity 2.5 Ah
Charging time 130 Minutes
Operating time 60 minutes with typical ventilation and ambient
temperature 25 °C (77 °F), see page 133
Remaining operating time At least 3 minutes (time between generation of Int.
battery low !!! alarm and failure of power supply)
Service life Depends on maintenance of internal battery
External battery 12 to 24 V DC (–10 to +30 %), 6.3 to 3.2 A
The optional external battery pack (EBP) is no lon-
ger available since 2014.
External battery pack (EBP) with internal
charger
Observe corresponding instructions for use.
The external battery pack (EBP) is a component of
the Carina medical system.
Gas supply
HPO
O
2 operating pressure 2.7 bar to 6 bar
270 kPa to 600 kPa
40 psi to 87 psi
O2 flow 0 to 120 L/min
O
2 connection Country-specific