Instructions for use Carina SW 3.2n 109
Reprocessing
Validated reprocessing procedures
The suitable reprocessing procedures result from
the classification.
The effectiveness of the reprocessing procedures
listed has been validated by independent laborato-
ries which are certified to ISO 17025.
At the time of validation, the following reprocessing
procedures were effective and compatible with the
materials used.
Reprocessing non-critical components
Surface disinfection with cleaning
1 Remove soiling immediately. To remove, use a
cloth moistened with disinfectant.
2 Perform surface disinfection.
3 After the contact time has elapsed, remove dis-
infectant residue.
4 Check the product for visible contamination.
Repeat steps 1 to 4 as required.
5 Check product for visible damage and replace it
if necessary.
Classification Procedure Agent Manufac-
turer
Concen-
tration
Contact
time
Tempera-
ture
Non-critical Surface disinfection
with cleaning
Virex TB Diversey Dosing and contact
time as specified by
manufacturer
–
Semi-critical A
and B
Machine cleaning
followed by thermal
disinfection
Neodisher
MediClean
Dr. Weigert Dosing and contact
time as specified by
manufacturer
–
–– –10min93°C
(199.4 °F)
Steam sterilization – – – 5 min 134 °C
(273.2 °F)
WARNING
Risk from penetrating liquid
Penetrating liquid my cause the following:
– Damage to the device
– Electric shock at switch-on
– Device malfunctions
Make sure that no liquid penetrates into the
device.