Getting started
58 Instructions for use Carina SW 3.2n
Safety information on getting started
Before getting started and after extended downtime
charge the internal battery of the device for at least
2 hours.
Requirements for getting started
– Device is reprocessed and assembled ready for
operation.
– Mains power supply or power supply with a
charged battery is established.
–O
2 supply is connected, if necessary.
Powering on the device
Set the power switch (A) to .
Carina runs the selftest. The screen shows the pro-
gress display for the selftest.
After the selftest the screen shows the last defined
ventilation mode and the last defined parameters.
The device is in standby mode. The patient is not
ventilated in standby mode.
WARNING
Risk of patient injury
Ventilation does not take place in standby
mode. Patients connected to the device are
endangered.
The device must only be set to standby mode
when no patient is connected to the device.
CAUTION
Malfunctions from condensation
Condensation may form when the device is
moved from a cold storage location to a warm en-
vironment.
Only switch on the device once the condensation
has dried.