Instructions for use Carina SW 3.2n 129
Technical data
Monitoring functions
Sound pressure level LPA of the alarm signals,
measured according to IEC60601-1-8 and
A1:2012:
Defined user's operating location: Front of device
at 1 m (39 in) distance and 1.5 m (59 in) height.
Acoustic alarm signal, standard setting (according
to IEC 60601-1-8)
Range for high-priority alarms approx. 73 dB (A) to approx. 78 dB (A)
Range for medium-priority alarms approx. 68 dB (A) to approx. 76 dB (A)
Range for low-priority alarms approx. 53 dB (A) to approx. 73 dB (A)
Step size Can be set in 5 steps
Acoustic alarm signal/ Dräger setting (common
acoustic alarm signal for Dräger ventilators)
Range for high-priority alarms approx. 67 dB (A) to approx. 81 dB (A)
Range for medium-priority alarms approx. 68 dB (A) to approx. 83 dB (A)
Range for low-priority alarms approx. 65 dB (A) to approx. 78 dB (A)
Step size Can be set in 5 steps
Power supply failure alarm and auxiliary alarm approx. 60 dB (A)
Inspiratory minute volume MV
Alarm for upper alarm limit If the upper alarm limit has been exceeded
Range 2 to 60 L/min
Resolution 0.1 L/min
Alarm for lower alarm limit If the lower alarm limit has been exceeded
Range 0.1 to 39 L/min
Resolution 0.1 L/min
Airway pressure Paw
Alarm for upper alarm limit If the alarm limit has been exceeded
Range 10 to 55 mbar (or hPa or cm H
2O)
Resolution 1 mbar (or hPa or cm H2O)
Delay time to alarm Airway pressure low !!! Tdiscon
Non-invasive ventilation (NIV) 0, 15, 30, 60, 90, 120 s
Invasive ventilation (Tube) 0, 15, 30, 60 s
Respiratory rate RR high
Alarm If the alarm limit has been exceeded for one minute
Range 10 to 74/min, Off
Resolution 1/min