Instructions for use Carina SW 3.2n 83
Alarms
Alarm priorities
The background color of the alarm line indicates
the priority of the active alarm. The priority of alarm
messages is also displayed by exclamation marks.
The alarm with the highest priority is signaled vis-
ually and acoustically.
For a list of causes and remedies, see "Alarm –
Cause – Remedy" on page 97.
Confirming alarms
The following alarms must be confirmed:
Airway pressure high !!!
Check settings !!
Invalid date !
To confirm, press the rotary knob.
Color of the
alarm line
Priority of the alarm message Behavior of the
alarm LED
Action required
Red back-
ground color
Warning !!! Alarm with high prior-
ity
Alarm LED
flashes red
Immediate action re-
quired to avert acute
danger
Yellow back-
ground color
Caution !! Alarm with medium
priority
Alarm LED
flashes yellow
Quick action required to
avert danger
Blue back-
ground color
Note ! Alarm with low prior-
ity
Alarm LED
lights yellow
Attention and action re-
quired

Table of Contents