Maintenance
112 Instructions for use Carina SW 3.2n
Overview
This chapter describes the maintenance measures
required to maintain the proper functioning of the
medical device. Maintenance measures must be
performed by the personnel responsible.
Only carry out maintenance work when no patient
is connected to the device.
Definition of maintenance concepts
CAUTION
Risk of infection
Users and service personnel can become in-
fected with pathogenic germs.
Disinfect and clean device or device parts before
any maintenance measures and also before re-
turning the medical device for repair.
WARNING
Risk of electric shock
Current-carrying components are located un-
der the cover.
Do not remove the cover.
Maintenance measures must be performed
by the personnel responsible. Dräger rec-
ommends DrägerService to perform these
measures.
CAUTION
Risk of electric shock
The batteries must be replaced by specialized
service personnel.
Concept Definition
Maintenance All measures (inspection, preventive maintenance, repair) intended to maintain
and restore the functional condition of a medical device
Inspection Measures intended to determine and assess the actual state of a medical device
Preventive mainte-
nance
Recurrent specified measures intended to maintain the functional condition of a
medical device
Repair Measures intended to restore the functional condition of a medical device after a
device malfunction