Maintenance
114 Instructions for use Carina SW 3.2n
Preventive maintenance
The following table shows the preventive mainte-
nance intervals:
Repair
Dräger recommends that all repairs are carried out
by DrägerService and that only authentic Dräger
repair parts are used.
WARNING
Risk of faulty components
Device failure is possible due to wear or mate-
rial fatigue of the components.
To maintain the proper operation of all compo-
nents, this device must undergo inspection
and preventive maintenance at specified inter-
vals.
Component Interval Measure Personnel responsi-
ble
Inlet filter (HEPA filter) Every 6 months or as
specified by hospital
regulations
Replace, see page 115 Users
Internal battery (NiMH
battery)
Every 3 years Replace Specialized service
personnel
Interior of device Every 3 years Visual inspection, clean and
replace affected parts if nec-
essary
Specialized service
personnel