Operation
72 Instructions for use Carina SW 3.2n
Canceling AutoAdapt
1 Press the (A) key three times. Menu 4 is set.
2 Press the key for Aut.Adapt (B). Select Off by
turning the rotary knob (F) and push to confirm.
The device ventilates with the previous ventilation
parameters.
Displaying setting and measured values
Prerequisite: Carina is in ventilation mode.
1 Press the Select Menu (A) key.
2 Press the Values (B) key.
The device shows the following on the screen:
Bar graph for pressure (C)
Set values (D)
Measured values (E)
Measured values depending on the valve in use
103104
B
A
C
DE
Valve Volume measured values
LeakV MVi, VTi, MVleak
1)
1) MVleak always shows the total leakage, consisting of pa-
tient leakage and valve leakage.
ExpV MV, VT, - -
2)
2) MVleak is always at - - because no leakage is measured
and compensated.