Instructions for use Carina SW 3.2n 85
Alarms
Exceeding the setting limit
The device limits specific alarm limits to a setting
limit.
1 Confirm the setting limit with the rotary knob.
Or
2 Set the parameter named in the message and
confirm with the rotary knob.
The set limit can be exceeded.
Switching off alarm limits
The following alarm limits can be switched off:
– RR high
– Tapn
To switch off the alarm limit
1 Touch the corresponding key for the alarm limit.
2 Set the highest value that can be set by turning
the rotary knob and push to confirm.
3 Turn the rotary knob further until Off is dis-
played and push to confirm.
The alarm limit is switched off.
Carina displays the symbol (A) in the alarm
line.
If an acoustic alarm signal is suppressed at the
same time, the following symbols are displayed al-
ternately:
– Symbol indicating switched off alarm limits
– Symbol indicating suppressed acoustic alarm
signals
WARNING
Risk of patient injury
When alarm limits are switched off, Carina
cannot monitor the patient.
Alarm limits may only be switched off if the
safety of the patient is not jeopardized by the
absence of an alarm.
212