Instructions for use Carina SW 3.2n 137
Technical data
EMC declaration
General information
The device was tested for electromagnetic compat-
ibility using accessories from the list of accessories.
Other accessories may only be used if they do not
compromise the electromagnetic compatibility. The
use of non-compliant accessories may result in in-
creased electromagnetic emissions or decreased
electromagnetic immunity of the device.
The device may be used in the direct vicinity of oth-
er devices only if Dräger has authorized this device
arrangement. If no authorization has been given by
Dräger, it must be ensured that the device functions
correctly in the desired arrangement before using it.
The instructions for use for the other devices must
be followed.
Electromagnetic environment
The emissions from this device were tested in the
following frequency ranges:
Electromagnetic immunity
This device may only be used in environments
specified in section "Intended use" on page 16.
Emissions Compliance
Conducted emissions Class B, group 1 (150 kHz to 30 MHz)
Radiated emissions Class B, group 1 (30 MHz to 1 GHz)
Harmonic emissions Requirement is met
Voltage fluctuation emissions (flicker) Requirement is met
Immunity against Test level
Electrostatic discharge (ESD) (IEC 61000-4-2) Contact discharge: ±8 kV
Air discharge: ±15 kV
Fast transient electrical disturbances (bursts)
(IEC 61000-4-4)
Power cable: ±2 kV
Longer input/output lines: ±1 kV
Impulse voltages (surges) (IEC 61000-4-5) Voltage, external conductor – external conductor:
±1 kV
Voltage, external conductor – protective ground con-
ductor: ±2 kV
Magnetic fields at a mains frequency of 50 Hz
(IEC 61000-4-8)
30 A/m