Assembly and preparation
34 Instructions for use Carina SW 3.2n
Safety information for preparation
Preparing Carina
Installing the inlet filter
1 Remove the inlet filter from the packaging.
2 Insert the inlet filter (A) into the holder from be-
low, fold it up and push it into the holder (B) until
it locks into position.
Positioning Carina
WARNING
Risk of personal injury
If medical devices are not reprocessed, there
is an increased risk of infection to both hospi-
tal staff and patients.
Before each use, reprocess the device and all
accessories in accordance with the instruc-
tions for use, see "Reprocessing list" on
page 110. Observe the hospital hygiene regu-
lations.
WARNING
Risk of damage to device or personal injury
If the device is not securely fastened, the de-
vice can fall down.
Securely fasten the device. Check for secure
fit.
023
WARNING
Risk of fire
Due to oxygen enrichment in the ambient air,
the medical device can ignite.
A distance of at least 10 cm (3.9 in) must be
maintained between the rear of the medical
device and the walls or large obstacles.
WARNING
Risk of overheating of the device
Pillows and blankets will block air feed.
Do not place the device on the patient's bed.