Technical data
138 Instructions for use Carina SW 3.2n
Recommended separation distances from mobile high-frequency communication
equipment
In order to attain the full functional integrity of this
device, a separation distance of at least 1.0 m
(3.3 ft) is required between the device and wireless
high-frequency communication equipment.
Voltage dips and short interruptions in the supply
voltage (IEC 61000-4-11)
30% to 100%, 10 ms to 5 s, different phase angles
Radiated high-frequency disturbances
(IEC 61000-4-3)
80 MHz to 2.7 GHz: 10 V/m
Conducted high-frequency disturbances
(IEC 61000-4-6)
150 kHz to 80 MHz: 3 V, ISM bands and amateur ra-
dio bands: 6 V
Electromagnetic fields in the vicinity of wireless
high-frequency communication equipment
Diverse frequencies from 385 MHz to 5785 MHz:
9 V/m to 28 V/m
WARNING
Risk due to insufficient separation distance
If mobile high-frequency communication
equipment is used too close to this device,
malfunctions that put the patient at risk may
occur.
Maintain a separation distance of at least
0.3 m (1.0 ft) between this device, including
the cables, and high-frequency communica-
tion equipment.
Immunity against Test level