Instructions for use Carina SW 3.2n 47
Assembly and preparation
Preparing Aeroneb Pro medication nebulizer
Observe the instructions for use of the Aeroneb
Pro nebulizer.
Observe the "Safety information on breathing
circuits and additional components"
on page 39.
WARNING
Increased resistance
If an HME and a medication nebulizer are used
at the same time, resistance can increase.
During medication nebulization, do not use a
HME.
WARNING
Risk of increased CO2 rebreathing
When using the breathing circuit with leakage
valve, aerosols may cause deposits in the
leakage valve. This will reduce the flow
through the leakage valve.
Use suitable external patient monitoring (cap-
nography) to detect increased CO
2 rebreath-
ing.
WARNING
Risk of increased CO2 rebreathing
Position the nebulizer between the inspiratory
hose and expiratory valve or leakage valve to
prevent increased dead space.
When using a medication nebulizer, use only
single hoses with a separate valve (MP00220,
MP00224).