Technical data
128 Instructions for use Carina SW 3.2n
Respiratory rate RR
Range 0 to 75/min
Resolution 1/min
Accuracy ±1/min
T
0...90 approx. 60 seconds
Mean airway pressure Pmean
Range 0 to 68 mbar (or hPa or cm H
2O)
Resolution 0.1 mbar (or hPa or cm H
2O)
Accuracy ±(8 % of measured value +1.2 mbar) when using
MP00312, MP00313 breathing circuits (without
leakages)
Minute volume MVi, MV
Range 0 to 60 L/min, BTPS
Resolution 0.1 L/min
Accuracy ±10 % of measured value or ±0.6 L/min, whichever
is greater (without leakage, and taking account of
hose compliance correction, "Volume losses during
ventilation" on page 144)
T
0...90 approx. 35 s
Leakage minute volume MVleak
Range 0 to 99 L/min, BTPS
Resolution 0.1 L/min
Accuracy ±10 % of measured value or ±0.6 L/min, whichever
is greater (without leakage, and taking account of
hose compliance correction, "Volume losses during
ventilation" on page 144)
T
0...90 Approx. 35 s
Adjustment time in AutoAdapt Tadpt
Range 0 to 120 min
Resolution 1 min
Accuracy 1 min
Airway pressure waveform display Pressure
Range 0 to 80 mbar (or hPa or cm H
2O)
Flow waveform display Flow
Range –160 to +160 L/min
Airway pressure bar graph Pressure
Range 0 to 60 mbar (or hPa or cm H
2O)
Displayed measured values (cont.)