Instructions for use Carina SW 3.2n 123
Technical data
Inspiratory pressure Pinsp
Only values from PEEP + 2 mbar can be set
Range (leakage valve) 5 to 40 mbar (or hPa or cm H
2O)
Range (expiratory valve) 5 to 50 mbar (or hPa or cm H
2O)
Resolution 1 mbar (or hPa or cm H
2O)
Pressure support
Δ
Psupp
Range (leakage valve) 0 to 37 mbar (or hPa or cm H
2O)
Condition: PEEP + ΔPsupp 40 mbar
(or hPa or cm H
2O)
Range (expiratory valve) 0 to 47 mbar (or hPa or cm H
2O)
Condition: PEEP + ΔPsupp 50 mbar
(or hPa or cm H
2O)
Resolution 1 mbar (or hPa or cm H
2O)
Respiratory rate RR
Range 5 to 50/min
Resolution 1/min
Inspiratory time Ti
Range 0.3 to 8 s
Resolution 0.1 s
Maximum inspiratory time for supported breaths Timax
Range 0.4 to 10 s
Resolution 0.1 s
Ratio of inspiratory time to expiratory time I:E
Range 1:39 to 2:1
Resolution 0.1
Pressure rise time Ramp
Range Auto (The device calculates the optimum value
from the measured respiratory rate)
0.1 to 2.0 s
Resolution 0.1 s
Trigger sensitivity Trigger
Range Off, Normal, Sensitiv.
Apnea ventilation Apn.Vent.
Range On, Off
Respiratory rate with apnea ventilation RRapn
Range 5 to 50/min
Resolution 1/min
Setting values (cont.)