Technical data
126 Instructions for use Carina SW 3.2n
2165619 breathing hose with 8412217 hu-
midifier chamber, 2 x 2165627 breathing
hose, 8404985 water trap and MP00220
expiratory valve
<4 mbar (or hPa or cm H
2O)
Expiratory resistance (adults) at 60 L/min
MP00312 breathing circuit <5 mbar (or hPa or cm H
2O)
MP00313 breathing circuit <5 mbar (or hPa or cm H
2O)
5704962 breathing circuit <5 mbar (or hPa or cm H
2O)
2166046 breathing circuit with MP00224
leakage valve
<5 mbar (or hPa or cm H
2O)
2166046 breathing circuit with MP00220
expiratory valve
<5 mbar (or hPa or cm H
2O)
F&P RT219 breathing circuit with MP00224
leakage valve
<5 mbar (or hPa or cm H
2O)
F&P RT219 breathing circuit with MP00220
expiratory valve
<5 mbar (or hPa or cm H
2O)
2165619 breathing hose with 8412217 hu-
midifier chamber, 2 x 2165627 breathing
hose, 8404985 water trap and MP00224
leakage valve
<5.5 mbar (or hPa or cm H
2O)
2165619 breathing hose with 8412217 hu-
midifier chamber, 2 x 2165627 breathing
hose, 8404985 water trap and MP00220
expiratory valve
<4 mbar (or hPa or cm H
2O)
Expiratory resistance (pediatric patients) at
30 L/min
MP00312 breathing circuit <2 mbar (or hPa or cm H
2O)
MP00313 breathing circuit <2.5 mbar (or hPa or cm H
2O)
5704962 breathing circuit <2.5 mbar (or hPa or cm H2O)
2166046 breathing circuit with MP00224
leakage valve
<2 mbar (or hPa or cm H
2O)
2166046 breathing circuit with MP00220
expiratory valve
<2.5 mbar (or hPa or cm H2O)
F&P RT219 breathing circuit with MP00224
leakage valve
<2 mbar (or hPa or cm H
2O)
F&P RT219 breathing circuit with MP00220
expiratory valve
<2.5 mbar (or hPa or cm H2O)
Performance characteristics (cont.)