EasyManuals Logo

Dräger Fabius Tiro User Manual

Dräger Fabius Tiro
224 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #103 background imageLoading...
Page #103 background image
Instructions for use Fabius Tiro SW 3.n 103
Operation
Adopting ventilation settings during
mode change
The ventilation settings for the new ventilation
mode are automatically derived from the settings
of the previous ventilation mode. The
corresponding settings (A) in the new ventilation
mode are highlighted.
The settings for Freq, TI:TE and PEEP are
adopted directly from the settings of the previous
ventilation mode if necessary.
When changing from Volume Control to
Pressure Control:
PINSP is set to the plateau pressure (PLAT)
that occurred in Volume Control.
Insp Flow is set to the last used value or the
factory setting.
When changing from Pressure Support to
Pressure Control:
Insp Flow is set to the last used value or the
factory setting.
When changing from Pressure Control to
Volume Control:
VT is set to the value that results from the
division of the last minute volume (MV) by the
respiratory rate (Freq).
TIP:TI is set to the last used value or the
factory setting.
PMAX is set to a value that lies 10 cmH
2O
(hPa) above the plateau pressure (PLAT) that
occurred in Pressure Control.
When changing from Volume Control to
Pressure Support:
Insp Flow is set to the last used value or the
factory setting.
PPS is set to the last used value or the
factory setting.
Trigger is set to the last used value or the
factory setting.
When changing from Pressure Control to
Pressure Support:
Insp Flow is set to the last used value or the
factory setting.
PPS is set to the last used value or the
factory setting.
Trigger is set to the last used value or the
factory setting.
When changing from Volume Control to SIMV/PS
PMAX and PEEP are automatically adopted for
the new ventilation mode from the previous
mode.
When changing from Pressure Support to
SIMV/PS:
PPS , Insp Flow , Trigger, and PEEP are
automatically adopted for the new ventilation
mode from the previous mode.
When changing from SIMV/PS with Pressure
Support to Pressure Support:
PPS and Insp Flow are automatically
adopted for the new ventilation mode from the
previous mode.
When changing from SIMV/PS to Pressure
Support:
Trigger and PEEP are automatically adopted
for the new ventilation mode from the previous
mode.
21725
Volume Control
A

Table of Contents

Other manuals for Dräger Fabius Tiro

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Fabius Tiro and is the answer not in the manual?

Dräger Fabius Tiro Specifications

General IconGeneral
BrandDräger
ModelFabius Tiro
CategoryMedical Equipment
LanguageEnglish

Related product manuals