Technical data
200 Instructions for use Fabius Tiro SW 3.n
Essential performance characteristics
The essential performance features comprise:
– Supplying the anesthesia workstation with O
2
If the O2 supply (central gas supply or gas
cylinder) fails, an alarm is issued.
– Supply of the patient with adequately
oxygenated breathing gas
If the breathing gas contains insufficient levels
of O
2, an alarm is issued.
– Monitoring of the airway pressure and the
expiratory minute volume
Alarms are issued depending on the set alarm
limits.
– Measurement accuracy of the O
2
measurement.
Alarms are issued depending on the set alarm
limits. If the O
2 sensor fails, an alarm is issued.
EMC declaration
General information
The EMC compliance of the product has been
evaluated with the external cables, transducers,
and accessories specified in the list of
accessories. Other accessories which do not affect
EMC compliance may be used if no other reasons
forbid their use (see other sections of the
instructions for use).The use of noncompliant
accessories can result in increased emissions or
decreased immunity of the medical device.
The medical device must only be used adjacent to
or stacked with other devices if this configuration is
approved by Dräger. If adjacent or stacked use of
configurations not approved by Dräger is
inevitable, verify correct functioning of the medical
device in this configuration before it is used. In any
case, strictly observe the instructions for use of the
other devices.
NOTE
In accordance with general safety standards,
additional components are required for a
complete anesthesia workstation.