Configuration
130 Instructions for use Fabius Tiro SW 3.n
Troubleshooting with failed test of the
overpressure safety valve
Repeat leakage test. If the test of the
overpressure safety valve continues to fail,
contact DrägerService or the authorized local
service partner.
Accessing the alarm logbook
The alarm logbook lists all alarm messages with
the respective date and time.
Up to a maximum of 100 entries can be saved.
When the storage limit is reached, the oldest
entries are overwritten.
1 Press the Access Alarm Log softkey (A).
2 To scroll through the alarm logbook (B), turn
the rotary knob.
Clearing the alarm logbook
Select Clear Alarm Log (C) and confirm
WARNING
Risk of unexpected occurring overpressure
A soiled or non-functioning overpressure
safety valve is not able to compensate for
suddenly occurring overpressure in the
breathing system.
Perform the leakage test before start-up of
the device. Observe the test results of the
overpressure safety valve.