Technical data
192 Instructions for use Fabius Tiro SW 3.n
Vaporizer interface
The anesthesia workstation is equipped with an interlock system.
When removing the vaporizer, the connection is automatically closed and sealed.
The following vaporizers can be used:
– Dräger-Vapor for halothane
– Dräger-Vapor for enflurane
– Dräger-Vapor for isoflurane
– Dräger-Vapor for sevoflurane
– Datex-Ohmeda Devapor/D-Tec for desflurane
– Dräger D-Vapor
Technical data of the vaporizers are contained in the corresponding instructions for use.
Measured value or
waveform
Range Resolution Accuracy Condition
PAW Airway pressure
(numeric)
-20 to 99 cmH
2O
(hPa)
1cmH2O
(hPa)
±4 %
1)
Airway pressure
(waveform)
0 to 99 cmH2O
(hPa)
Pressure gauge
(mechanical)
-20 to 80 cmH
2O
(hPa)
2cmH2O
(hPa)
1.28 cmH2O (hPa)
MVe Expiratory minute
volume
0 to 32.0 L/min 0.1 L/min ±15% or ±0.2 mL,
depending on which
value is higher
2)
Based on 20 °C
(68 °F) Ambient
pressure and satu-
rated gas
VTe Expiratory tidal
volume
0 to 1500mL 1mL ±15%
2)
or ±20 mL,
depending on which
value is higher

Table of Contents

Other manuals for Dräger Fabius Tiro

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Fabius Tiro and is the answer not in the manual?

Dräger Fabius Tiro Specifications

General IconGeneral
BrandDräger
ModelFabius Tiro
CategoryMedical Equipment
LanguageEnglish

Related product manuals