Instructions for use Fabius Tiro SW 3.n 49
Operating concept
Total flow tube
The total flow tube is used as a reference for the
total volume of fresh gas that is introduced in the
breathing system. The individual measured flow
values for N
2O, Air, and O2 are displayed on the
respective electronic fresh-gas flow displays.
The total flow tube remains functional if the
following faults occur:
Fault in the electronic flow measurement
Fault in the digital display
Fault in the current switching circuit
In these cases, the measurement of the total flow
tube indicates the total flow measured before the
fault occurred.
To adapt the fresh-gas ratio during an existing
fault, close all flow control valves and reset the
individual flows one after another. The O
2 valve
can remain open.
Example: After closing the flow control valves, the
total flow tube indicates 2 L/min. This corresponds
to an O
2 flow of 2 L/min.
If 1 L/min N
2O is required, open the N2O flow
control valve until the total flow tube displays
3L/min.
Monitoring resolutions for fresh-gas
flow
Fabius can be configured for the following
resolutions of fresh-gas flow by DrägerService or
the authorized local service partner:
Standard resolution (A)
High resolution (B)
If a flow is higher than 9.99 L/min, the standard
resolution is activated.
If the standard resolution is configured, the
electronic fresh-gas flow displays show the flow in
increments of 100 mL/min (format xx.x L/min). The
virtual flow tubes (A) of the screen indicate a range
from 0 to 12 L/min.
If a high resolution is configured, the electronic
fresh-gas flow displays show the flow in
increments of 10 mL/min (format x.xx L/min). The
virtual flow tubes (B) of the screen indicate a range
from 0 to 10 L/min.
If the respective flow again drops below
9.00 L/min, the high resolution is activated.
NOTE
The total flow tube is calibrated for a mixture of
N2O and O2 in a 50/50 ratio. The accuracy of the
total flow tube might be reduced with other gas
mixtures. More information can be found in
chapter ''Technical data'' on page 182.
2158221587
A
Standby
Standby
B