Troubleshooting
152 Instructions for use Fabius Tiro SW 3.n
Alarm – Cause – Remedy
Alarm messages are displayed in hierarchical form
in the alarm message field of the main screen, see
chapter ''Screen display'' on page 45.
The priority of the alarm messages is marked by
exclamation points.
The alarm messages are only displayed on
colored background if the option "Color display" is
activated.
Within an alarm priority, the alarm messages are
assigned internal priorities. In the following table,
these internal priorities are indicated as numbers.
The alarm message with the highest priority has
the number 31. The lower the priority, the lower the
number.
The table lists possible causes and the
corresponding remedial measures of an alarm.
Causes and remedial measures must be
processed in the sequence they are listed until the
alarm no longer occurs. The alarm messages are
listed in alphabetical order.
Some alarms appear in this table several times
with different alarm priorities because their priority
can change under certain conditions.
Warning !!! Red
Caution !! Yellow
Note ! White
Alarm
priority
Alarm Cause Remedy
(31) AIRWAY PRESSURE HIGH !!! The upper alarm limit for the
airway pressure was
exceeded, the breathing hose
is kinked.
Check the breathing circuit
connected to the anesthe-
sia workstation.
(31) AIRWAY PRESSURE HIGH !!! The alarm limit was set too
low.
Check the breathing sys-
tem or the alarm limit.