Assembly and preparation
68 Instructions for use Fabius Tiro SW 3.n
Connecting the exhaust port
Screw on the exhaust port (A) in the compact
breathing system from below.
Connecting the breathing bag
The breathing bag can be mounted to the compact
breathing system with the following variants:
– On a flexible arm
– On a rigid arm
– On the bag elbow directly on the compact
breathing system
20968
WARNING
Risk of use of toxic or incompatible materials
The breathing bag used must comply with the
current standards.
WARNING
Risk of too high airway pressure or
insufficient fresh-gas
If the breathing bag is pinched, excessive
airway pressures or a lack of fresh gas can
occur.
Attach and position the breathing bag so that
it is not pinched and can inflate freely.
CAUTION
Risk of patient recovering consciousness
A blocked or incorrectly positioned breathing bag
can lead to lack of fresh gas for patients. Manual
ventilation is also not possible.
Make sure that the breathing bag is connected
tightly to the bag holder during the ventilation.