Instructions for use Fabius Tiro SW 3.n 69
Assembly and preparation
Mounting the flexible arm
Prerequisite: Before mounting the flexible
breathing bag holder, the bag elbow on the
compact breathing system must be removed.
1 Position and align the connection adapter (A)
of the flexible arm on the connection of the
compact breathing system.
2 Tighten the knurled screws (B). Check that the
arm is fixed securely.
3 Fasten the elbow (D) at the end of the flexible
arm.
4 Fasten the breathing bag (C) on the other end
of the elbow.
Mounting the rigid arm
Prerequisite: Before mounting the rigid breathing
bag holder, the bag elbow on the compact
breathing system must be removed.
1 Position and align the connection adapter (A)
of the rigid arm on the connection of the
compact breathing system.
2 Tighten the knurled screws (B). Check that the
arm is fixed securely.
3 Fasten the breathing bag (C) at the other end
of the rigid arm.
20975