Alarms
114 Instructions for use Fabius Tiro SW 3.n
Alarm priorities
Fabius assigns the appropriate priority to each
alarm.
The background color of the alarm message field
indicates the priority of the active alarm.
The alarm messages are only displayed on
colored background if the option "Color display" is
activated.
The alarm messages are sorted according to these
priorities and displayed corresponding to the
internal priority system. Priority 31 has the highest
and priority 1 the lowest priority. The priority
numbers are listed in the table in chapter ''Alarm –
Cause – Remedy'' on page 152.
There can be a maximum of 4 alarm messages
*
displayed at the same time in a list. Alarm
messages with higher priority are displayed before
alarm messages with lower priority. Alarm
messages with low priority are only displayed if the
cause for a high-priority alarm is remedied.
Color Priority of the alarm message Required action
Red Warning High-priority alarm !!! Immediate action is necessary in order to
avert imminent danger.
Yellow Caution Medium-priority alarm !! Fast action is necessary in order to avert a
danger.
Note Low-priority alarm ! Attention is necessary, but a delayed
response is sufficient.
* optional 3 alarm messages for Japan and China