Instructions for use Fabius Tiro SW 3.n 189
Technical data
Ventilator
ISO 80601-2-55 Part 2-55:
Particular requirements for basic safety and essen-
tial performance of respiratory gas monitors
In compliance with ISO 80601-2-13
Control ranges
Pressure limitation (PMAX) 15 to 70 cmH
2O (resolution: 1 cmH2O)
(15 to 70 hPa (resolution: 1 hPa))
(setting must be at least 10 cmH
2O (10 hPa)
above PEEP, in SIMV/PS mode, the PMAX set-
ting must be greater than ∆PPS+PEEP)
Tidal volume (VT) 20 to 1400 mL (resolution: 10 mL)
Tidal volume (VT) 20 to 1100 mL (resolution: 10 mL), in SIMV/PS
mode
Respiratory rate (Freq) 4 to 60 bpm (resolution: 1 bpm)
(4 to 60 1/min (resolution: 1/min))
Ratio of inspiratory time to expiratory time (TI:TE) 4:1 to 1:4
Inspiratory pause (TIP:TI) 0 % to 50 % (resolution: 1%)
Positive end-expiratory pressure (PEEP) 0 to 20 cmH
2O (resolution: 1 cmH2O)
(0 to 20 hPa (resolution: 1 hPa))
Inspiratory pressure (PINSP) 5 to 65 cmH
2O (resolution: 1 cmH2O)
(5 to 65 hPa (resolution: 1 hPa)) (setting must be
at least 5 cmH
2O (5 hPa) above PEEP)
Inspiratory flow (Insp Flow) 10 to 75 L/min (resolution: 1 L/min) in Pressure
Control mode
10 to 85 L/min (resolution: 1 L/min) in Pressure
Support and SIMV/PS modes
Support pressure (∆PPS) 3 to 20 cmH
2O (resolution: 1 cmH2O)
(3 to 20 hPa (resolution: 1 hPa)), in Pressure
Support mode
Support pressure (∆PPS) 3 to 20 cmH
2O, OFF (resolution: 1 cmH2O)
(3 to 20 hPa OFF (resolution: 1 hPa)), in
SIMV/PS mode
Relevant standards (continued)