Instructions for use Fabius Tiro SW 3.n 177
Maintenance
Service
The following table shows the service intervals:
WARNING
Risk of faulty components
Device failure is possible due to wear or
material fatigue of the components.
To maintain the function of all components,
this device must be inspected and serviced at
the intervals specified by the manufacturer.
WARNING
Risk of electric shock
Before performing any service work,
disconnect all electrical connections and gas
connections from power and gas supplies.
Component Interval Measure Responsible personnel
CO
2 absorber If colored violet Replace Users
Water trap As needed or if
soiled
Replace Users
Flow sensor As needed or if
calibration is no
longer possible
Clean/replace Users
Internal lithium battery Every 36 months Replace Experts
Fabius Tiro Every 6 months Inspection and service Experts
Breathing system Every 6 months Inspection and service Service personnel
Vaporizers Every 6 months Inspection and service Service personnel
Sensors Every 6 months Inspection and service Service personnel
Lead-gel battery Every 3 years Replace Experts
Cylinder pressure reducer for
high-pressure cylinders
1)
1) optional
After 6 years Basic overhauling Experts
Pressure reducer for PIN
index
1)
After 6 years Replace Experts