Instructions for use Fabius Tiro SW 3.n 85
Getting started
Checking the readiness for operation
At the end of the system test, one of 3 possible
results (A) are displayed on the screen:
FUNCTIONAL
CONDITIONALLY FUNCTIONAL
NON-FUNCTIONAL
FUNCTIONAL
The device is ready for operation. After a short
delay, the page Standby is displayed.
CONDITIONALLY FUNCTIONAL
A non-critical malfunction was found. The
anesthesia workstation can be used.
To open the page Standby, press the rotary
knob.
Check if an alarm message is displayed.
Contact DrägerService or the authorized local
service partner.
NON-FUNCTIONAL
A serious malfunction was found and the operation
of the monitoring functions and ventilation
functions are blocked.
Do not use the device.
Immediately contact DrägerService or the
authorized local service partner.