Configuration
128 Instructions for use Fabius Tiro SW 3.n
Troubleshooting with failed O2 calibration
Replace the O
2 sensor capsule in the O2
sensor housing, see chapter ''Inserting a new
O
2 sensor capsule'' on page 75.
If the calibration continues to fail, contact
DrägerService or the authorized local service
partner.
Leakage test
The following tests are initiated during the leakage
test:
– Compliance Test
– System Leak Test
– Ventilator Leak Test
– Safety Relief Valves Test
1 Press the Leak /ComplTest softkey (A).
2 Follow the instructions on the screen.
After completion of the tests, the results (B) are
displayed on the screen.
3 To return to the start screen, press the rotary
knob.
Results of the compliance test
This test determines the system compliance
including breathing system, breathing hoses, filter,
and Y-piece.
The system compliance is required to ensure in
Volume Control mode that the applied tidal
volume corresponds to the set tidal volume.
The value of the compliance is displayed on the
Standby screen.
21582
21984
System compliance
[mL/cmH2O]
Displayed result
[mL/cmH2O]
≤6.5 Measured value and
PASSED