Instructions for use Fabius Tiro SW 3.n 129
Configuration
Results of the ventilator leakage test
The test of the ventilator leakage can have the
following results:
Results of the system leakage test
The test of the system leakage can have the
following results:
Test of the overpressure safety valve
This test checks the functionality of the
overpressure safety valve.
The test results are displayed on the screen with
the leakage test results (B).
Ventilator leakage
[mL/min]
Displayed result
[mL/min]
150 Measured value and
PASSED
151 to 250 Measured value and
FAILED
>250 >250 and FAILED