Technical data
202 Instructions for use Fabius Tiro SW 3.n
Electromagnetic environment
The medical device is intended for use in an
electromagnetic environment as specified below.
The user must ensure that the medical device is
used in such an environment.
Emissions Compliance according to Electromagnetic environment
Radio frequency emissions
(CISPR 11)
Group 1 The medical device uses radio
frequency energy only for its
internal function. Therefore, its
radio frequency emissions are
very low and are not likely to
cause any interference in nearby
electronic equipment.
Class A The medical device is suitable for
use in all establishments other
than domestic establishments
and those directly connected
(without transformer) to the public
low-voltage power supply net-
work that supplies buildings used
for domestic purposes.
Harmonic emissions (IEC 61000-
3-2)
Not applicable
Voltage fluctuations/flicker emis-
sions (IEC 61000-3-3)
Not applicable