Assembly and preparation
54 Instructions for use Fabius Tiro SW 3.n
1 Connect the power cable with the device.
2 Plug the power cable into the mains power
socket on the wall.
LED on the front of the device lights up
green.
3 Set On/Off switch (A) to position .
4 Check the status bar for the battery indicator.
Auxiliary power sockets
Establishing potential equalization
Differences in electrical potential between devices
can be reduced by potential equalization.
Potential equalization does not replace the
protective ground connection.
In operation, the potential equalization connectors
must be readily accessible and must be removable
without tools.
NOTE
The mains plug must be freely accessible so that
the power supply to Fabius can be quickly
interrupted in the event of device failure.
WARNING
Risk of electric shock
The connection of devices to auxiliary power
sockets can lead to an increased leakage
current. If the protective ground of one of
these devices fails, the leakage current may
rise above the permissible values.
Only connect with the approval of the
respective device manufacturer.
Have the leakage current checked by
service personnel.
If the permissible value is exceeded, use a
mains power socket on a wall instead of
the auxiliary power socket of the device.
WARNING
Risk of device malfunction
If the mains power fails, devices connected to
the auxiliary power sockets are not supplied
from the uninterruptible power supply.
Do not connect any life-supporting
devices to the auxiliary power sockets of
the anesthesia workstation.
Ensure an alternative power supply for
connected devices.
WARNING
Risk of device malfunction
If high-frequency surgical devices are
connected to the auxiliary power sockets, the
leakage current can damage the electronics
of the medical device and lead to a failure.
Do not connect any high-frequency surgical
devices to the auxiliary power sockets of the
medical device.

Table of Contents

Other manuals for Dräger Fabius Tiro

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Fabius Tiro and is the answer not in the manual?

Dräger Fabius Tiro Specifications

General IconGeneral
BrandDräger
ModelFabius Tiro
CategoryMedical Equipment
LanguageEnglish

Related product manuals