Cleaning, disinfection and sterilization
172 Instructions for use Fabius Tiro SW 3.n
Semi-critical medical devices
Components which
can be reprocessed
Recommended
reprocessing
intervals
Prelimi-
nary
cleaning
Machine
cleaning and
disinfection
Manual Steriliza-
tion
Cleaning Disinfec-
tion
Breathing system:
Breathing system
housing
Inspiratory/expira-
tory ports, APL
valve
Inspiratory valve,
expiratory valve
Bag elbow
Rigid arm for
breathing bag
(optional)
Flexible arm for
breathing bag
(optional)
Weekly Yes Yes Possible Possible Possible
Breathing hoses After each patient Observe the associated instructions for use.
Absorber container
and absorber insert
Weekly Yes Yes Possible Possible Possible
Soda lime dust filter
(optional)
Replace each
time soda lime is
changed.
No No No No No
Sample line Replacement
only
No No No No No
When the sample
line is fitted to the
filter on the Y-
piece.
Daily
When the sample
line is fitted directly
to the Y-piece and
the filters are fit-
ted on the breath-
ing system
After each patient
Ventilator lid After each
patient
Yes Yes No Yes Yes
Ventilator membrane After each
patient
Yes Yes No Yes Yes

Table of Contents

Other manuals for Dräger Fabius Tiro

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Fabius Tiro and is the answer not in the manual?

Dräger Fabius Tiro Specifications

General IconGeneral
BrandDräger
ModelFabius Tiro
CategoryMedical Equipment
LanguageEnglish

Related product manuals