EasyManua.ls Logo

Dräger Fabius Tiro - Wall-Mounted Version (Optional)

Dräger Fabius Tiro
224 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instructions for use Fabius Tiro SW 3.n 29
Overview
Wall-mounted version (optional)
Fabius can also be mounted on the wall. In this
case, the anesthesia workstation is not placed on a
trolley, (see chapter ''Fabius Tiro as trolley version
(front view)'' on page 20), but rather on a mount on
the wall.
Observe the corresponding assembly instructions
(9037202).
21008
CAUTION
Risk of tipping over
If the device is removed from the mount and set
on the floor, it can tip over. Observe the following
steps before setting the device on the floor:
All additionally mounted components must be
removed beforehand.
Fold the holding arms in on the Fabius.
Make sure there is sufficient free space.
CAUTION
Risk of injury or risk of damage to the medical
device
If the maximum weight is exceeded, the holder
can release from the wall mount.
The total weight of the anesthesia workstation
including all accessories must not be exceeded,
see ''Instructions for mounting the accessories''
and ''Technical data''.
CAUTION
Risk of injury or risk of damage to the medical
device
When the device is being mounted to the wall,
pivoted in its holder, or removed from the wall,
the cables and hoses may be damaged and
personal damage may occur.
Ensure that cables and hoses are not
pinched, kinked, or torn.
When mounting, removing, or pivoting the
device, ensure that body parts are not
crushed.
The device may only be moved by people
who have the physical ability to do so. Dräger
recommends that such tasks be carried out
by two people.

Table of Contents

Other manuals for Dräger Fabius Tiro

Related product manuals