Instructions for use Fabius Tiro SW 3.n 107
Operation
4 If necessary, connect the anesthetic gas
scavenging hose of the non-rebreathing
system to the second connection (D) of the
anesthetic gas receiving system.
Observe the instructions for use of the non-
rebreathing system and the anesthetic gas
receiving system.
Operation
1 Change to Standby mode.
2 Set the fresh-gas flow.
To prevent rebreathing, the fresh-gas supply
must be at least double the minute volume.
3 Operate the non-rebreathing system according
to the corresponding instructions for use.
Ending operation
1 Close all flow control valves on the device.
2 Disconnect the non-rebreathing system from
the external fresh-gas outlet.
3 Connect the fresh-gas hose to the external
fresh-gas outlet (A).
4 Screw the sample line back to the Y-piece on
the breathing circuit.
21020
D
21000
A