Technical data
198 Instructions for use Fabius Tiro SW 3.n
S-ORC (Sensitive Oxygen Ratio Controller)
At a flow of approx. 200 mL Set the N2O concentration in the fresh gas
between 0 and 75 %.
In case of insufficient O
2 S-ORC limits the N2O concentration in the fresh
gas so that the O
2 concentration does not drop
below 23 Vol%.
N
2O flow control valve is open and at the same
time the O
2 flow control valve is closed or set at
less than 0.2 L/min.
S-ORC prevents N
2O flow.
In case of N2O failure O2 can continue to be supplied. No alarm.