Dräger Fabius Tiro - S-ORC (Sensitive Oxygen Ratio Controller)

Dräger Fabius Tiro
224 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Technical data
198 Instructions for use Fabius Tiro SW 3.n
S-ORC (Sensitive Oxygen Ratio Controller)
At a flow of approx. 200 mL Set the N2O concentration in the fresh gas
between 0 and 75 %.
In case of insufficient O
2 S-ORC limits the N2O concentration in the fresh
gas so that the O
2 concentration does not drop
below 23 Vol%.
N
2O flow control valve is open and at the same
time the O
2 flow control valve is closed or set at
less than 0.2 L/min.
S-ORC prevents N
2O flow.
In case of N2O failure O2 can continue to be supplied. No alarm.