Instructions for use Fabius Tiro SW 3.n 187
Technical data
External fresh-gas outlet
Electrical safety
General safety standards for anesthesia workstations
Connection 22 mm outer taper / 15 mm inner taper (ISO)
Pressure limitation Max. 80 cmH
2O (hPa) at 18 L/min
Fresh-gas flow 0 and 0.2 to 18 L/min
In compliance with UL 60601-1
IEC 60601-1
CAN/CSA C22.2 No. 601.1-M90
Relevant standards
In addition to the standards listed here, this medical
device meets various other standards, e.g., stan-
dards concerning special national requirements.
IEC 60601-1 2nd ed.
Medical electrical equipment
Part 1:
General requirements for safety
IEC 60601-1-2
Medical electrical equipment
Part 1-2:
General requirements for safety, collateral stan-
dard: Electromagnetic compatibility – Requirements
and tests
IEC 60601-1-4
Medical electrical equipment
Part 1-4
General requirements for safety, collateral stan-
dard: Programmable electrical medical systems
IEC 60601-1-8
Medical electrical equipment
Part 1-8:
General requirements for safety, collateral stan-
dard: General requirements, tests, and guidance for
alarm systems in medical electrical systems