Assembly and preparation
60 Instructions for use Fabius Tiro SW 3.n
Mounting the anesthetic vaporizers
Depending on the configuration, Fabius can be
operated with vaporizers that have a plug-in
adapter for Selectatec or Interlock S compatible
connectors. The anesthetic vaporizers must be
mounted in accordance with the respective
instructions for use.
The vaporizers in use must comply with applicable
safety standards.
If an independent gas measurement system is
used, it must comply with applicable safety
standards.
WARNING
Risk due to incorrect anesthetic agent
delivery
If the vaporizer is filled with the wrong
anesthetic agent or if it is not filled
sufficiently, incorrect anesthetic gas
concentrations or concentrations that are too
low can occur as a result.
Strictly observe the instructions for use of
the vaporizer.
Compare the color coding on the
vaporizer with the anesthetic agent bottle.
WARNING
Risk due to improperly mounted vaporizers
Incorrectly mounted vaporizers can cause
leakage. This can cause the fresh-gas flow to
be too low or contaminate the ambient air.
Patient and user can be endangered.
Make sure that the vaporizers are
mounted levelly.
When using D-Vapor vaporizers, make
sure that the power cable is not pinched.
After mounting the vaporizers, perform a
leakage test.

Table of Contents

Other manuals for Dräger Fabius Tiro

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Fabius Tiro and is the answer not in the manual?

Dräger Fabius Tiro Specifications

General IconGeneral
BrandDräger
ModelFabius Tiro
CategoryMedical Equipment
LanguageEnglish

Related product manuals