Instructions for use Fabius Tiro SW 3.n 95
Operation
Ventilation mode Volume Control
Compensation of ventilator compliance
Ventilator compliance compensation is active in
Volume Control mode so that the tidal volume
administered to the patient (VT) corresponds to the
tidal volume setting. The ventilator compliance is
determined during the leakage test in Standby
mode, see chapter ''Leakage test'' on page 128.
The breathing hoses used during the leakage test
and compliance test must also be used during
operation.
This guarantees an exact compliance
compensation.
Changing to Volume Control mode
In the following examples and illustrations, the
change from Pressure Control mode to Volume
Control mode is described:
1 Press the Volume Control (A) key.
2 Adjust ventilation settings (B).
3 Confirm the new mode.
In the following table, all parameters (B) in Volume
Control mode, together with the corresponding
setting ranges and factory settings, are listed.
NOTE
If the ventilator works at its performance limit due
to the volume control settings, Fabius cannot
make the compliance compensation. If the
performance limit of the ventilator is reached, it is
not possible to increase the setting for tidal
volume VT.
21723
Pressure Control
A
B
21577
Parameter Setting range Factory set-
ting
Pressure limita-
tion PMAX
[cmH
2O]
([hPa])
15 to 70,
min. PEEP +10
40
Tidal volume
VT [mL]
20 to 1400 600
Respiratory
rate Freq [bpm]
([1/min])
4 to 60 12
Inspiratory
time:expiratory
timeTI:TE
4:1 to 1:4 1:2
Inspiratory
pause
time:inspira-
tory time TIP:TI
[%]
0 to 50 10
PEEP [cmH
2O]
([hPa])
0 to 20 0
Volume Control