Operation
94 Instructions for use Fabius Tiro SW 3.n
The following alarms can be activated or
deactivated in ManSpont mode with the ON/OFF
key (B):
Apnoea Pressure (see chapter ''Pressure
alarms in automatic ventilation modes'' on
page 122)
Volume Alarms (see chapter ''Volume
alarms'' on page 121)
4 To refill the breathing bag, press the key.
5 Set suitable fresh-gas flow.
6 Start manual ventilation with the breathing bag.
The pressure is limited to the value that is set
on the APL valve.
Pressure release
In the ManSpont mode, lifting the valve head
relieves pressure from the breathing system.