EasyManua.ls Logo

Dräger Fabius Tiro - Main Screen; O2 Monitoring

Dräger Fabius Tiro
224 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instructions for use Fabius Tiro SW 3.n 117
Monitoring
Main screen
Press the (A) key.
The current dialog window changes to the main
screen.
The following information is displayed on the
screen:
Current alarm messages
Data of the O
2 monitoring
Data of the airway pressure monitoring
Data of the breathing volume monitoring
O2 monitoring
The inspiratory oxygen concentration is measured
with a dual galvanic sensor, which is located in the
cover of the inspiratory valve, see chapter
''Compact breathing system COSY (top view)'' on
page 21.
Parameter field for O2 monitoring
The following information is displayed in the
parameter field for O2 monitoring:
A Numerical value for the inspiratory O
2
concentration in percent (%) in the range from
10 % to 100 %
B Upper alarm limit for the O
2 concentration
in (%)
C Lower alarm limit for the O
2 concentration
in (%)
21590
A
CAUTION
Risk of inaccurate measured values
If the O
2 sensor is removed, this can lead to
leaks in the breathing system.
When the O2 sensor is replaced or removed, it
must be recalibrated.
NOTE
If the anesthesia workstation is not used, remove
the O
2 sensor from the cover of the inspiratory
valve and insert the sealing plug provided.
21780
Volume Control
A
B
C

Table of Contents

Other manuals for Dräger Fabius Tiro

Related product manuals