Instructions for use Fabius Tiro SW 3.n 131
Configuration
Closing the alarm logbook
Select the input arrow (D) and confirm.
The screen changes to Standby mode.
Restoring the default settings
The default settings are restored in the following
cases:
Switching Fabius on and off
Performing the system test
Pressing the softkey Restore Site Defaults
1 Press the Restore Site Defaults softkey (A).
The default settings are restored. The following
message is displayed using the standby softkeys
(B):
Site Default settings restored
The default settings can be adjusted on the
Standby Set-up screen. Adjustment of the default
settings is password-protected.
CAUTION
Risk of losing data
All data in the alarm logbook are cleared in the
following cases:
Fabius is switched off.
The system test is started in standby mode.
The power supply fails.
21582
Standby
A
B
WARNING
Risk due to unsuitable ventilation settings
After the default settings are restored, check
whether the settings for the ventilation and
monitoring are suitable for the patient.