Configuration
144 Instructions for use Fabius Tiro SW 3.n
Changing the alarm volume
1 Press the AccessAlarm Volume softkey (A).
2 Set the new alarm volume to a value between 1
(minimum) and 10 (maximum) and confirm.
The lower value is limited by the setting in the
standby configuration (see chapter ''Changing the
minimum alarm volume'' on page 134).
The pressure waveform and softkeys are
displayed again.
WARNING
Risk of not hearing the alarm tone
When operating in a loud environment, the
acoustic alarm signals may not be heard.
Always set the alarm tone to a sufficient
volume.
2215522163