Instructions for use Fabius Tiro SW 3.n 191
Technical data
Anesthetic gas supply module
Fresh-gas flow indicators
O
2, N2O, Air Range and accuracy:
0.0 to 12.0 L/min ±10 % of the measured value or
±0.12 L/min, depending which value is higher,
against atmospheric pressure of 14.7 psi
(1.013 kPa x 100) at 20 °C (68 °F).
Resolution: 0.1 L/min
Fresh-gas flow stability
O
2 and N2O: ±10% of the setting at supply pressures of 41 to 87 psi (2.8 to 6 kPa x 100)
Air: ±10% of the setting at supply pressures of 50 to 55 psi (3.4 to 3.8 kPa x 100).
Outside the range of 50 to 55 psi (3.4 to 3.8 kPa x 100), the Air flow varies proportionally to the supply
pressures.
Total flow tube
Range and accuracy 0 to 10 L/min ±10 % of the measurement range at
standard temperature and standard pressure, cal-
ibrated with a gas mixture of 50% O
2 and
50% N2O
0 to 10 L/min ±15% of the measurement range at
standard temperature and standard pressure for
all other gas mixtures
Resolution 0.5 L/min at 0.5 to 2 L/min 1.0 L/min at
2to10L/min
O
2 flush at 87 psi (6 kPa x 100): max. 75 L/min
at 41 psi (2.8 kPa x 100): min. 25 L/min
Pressure limit of the common gas outlet Maximum 13 psi (0.9 kPa x 100) ±5%
Flow tube for O
2 supplemental delivery (optional)
Connection Stepped connection for use with hoses of different
diameters
Flow 0 to 10 L/min
Accuracy ±5 % of the measurement range
Resolution 0.5 L/min