Technical data
190 Instructions for use Fabius Tiro SW 3.n
Minimum respiratory rate for apnea ventilation
(Freq Min)
3 to 20 bpm (resolution: 1 bpm) and OFF
(3 to 20 1/min (resolution: 1/min) and OFF)
Trigger value (Trigger) 2 to 15 L/min (resolution: 1 L/min)
Inspiratory time (TINSP) 0.3 bis 4.0 sec.
Accuracy
Pressure limitation (PMAX5cmH2O (±5 hPa) of the setting
Tidal volume (VT) ±5 % of the setting or 20 mL, depending on which
value is higher (discharged to atmosphere, no
compliance correction)
Respiratory rate (Freq) ±1 bpm (±1 1/min) of the setting or ±5 %, depend-
ing on which value is higher
Ratio of inspiratory time to expiratory time (TI:TE) ±5 % of the setting
Inspiratory pause (TIP:TI) ±25 % of the setting
Positive end-expiratory pressure (PEEP2cmH
2O (±2 hPa) or ±20 % of the setting,
depending on which value is higher
Inspiratory pressure (PINSP2cmH
2O (±2 hPa) or ±20 % of the setting,
depending on which value is higher
Overpressure safety valve 75 ±5 cmH2O (75 ±5 hPa)
Underpressure safety valve (inlet valve for ambi-
ent air)
-7.5 to -9 cmH
2O (-7.5 to -9 hPa)
Minimum pressure limit -9 cmH
2O (-9 hPa)
Measuring the system compliance 0.2 to 6.0 mL/cmH
2O (0.2 to 6.0 mL/hPa)
±0.2 mL/cmH
2O (±0.2 mL/hPa) or ±10 % of the
actual compliance depending which value is
higher