Instructions for use Fabius Tiro SW 3.n 27
Overview
Vaporizer
Vaporizers are used to enrich fresh gas with a
precise delivered concentration of a volatile
anesthetic agent.
Vapor 2000/3000 is an unheated, calibrated
vaporizer for enriching dry, medical fresh gas of an
anesthesia workstation with a precise delivered
concentration of a volatile anesthetic agent.
D-Vapor/D-Vapor 3000 is a heated, calibrated
vaporizer for enriching dry, medical fresh gas of an
anesthesia workstation with the anesthetic agent
desflurane.
There are various connector systems with which
vaporizes can be connected to an anesthesia
workstation.
Dräger recommends using only vaporizers that are
listed in the list of accessories.
More information can be found in the respective
instructions for use of the anesthetic vaporizers
used.
20988
Vaporizer Anesthetic agent
Vapor 2000/3000
Isoflurane
Halothane
Enflurane
Sevoflurane
D-Vapor/ D-Vapor 3000 Desflurane